Quality Control of temperature monitoring
A New England blood banker reports that his facility uses a REES
temperature monitoring system to monitor all product storage and manufacturing
areas. According to the inquiring blood banker, the Code
of Federal Regulations 21;606.60(b), Revised as of April 1, 2001 applies to this type of monitoring system. However, he would like input from the e-Network as to the proper frequency that such a system should be validated/calibrated.
The following response was
received.
ADDENDA Dec. 17, 2001
- According to a responding blood banker, in his experience the REES system must be validated on site by each facility using it.
The user should validate each probe by the 'fire & ice' method '' using a calibrated thermometer when the system is installed and whenever changes are made (i.e., changing out a probe, sensor, or changing the set points for high and low readings). One must be able to validate that each sensor will do exactly what it was meant to do. Validation records must be kept on file for review by FDA. The responding blood banker states that the REES System probes are capable of being calibrated to within 0.2C of the alarm points. The responding blood banker reported that his facility routinely re-calibrates the probes at least once a year routinely, and each time they changed out a probe.
ADDENDA Jan. 28, 2002
- A blood banker from a community blood collection center in Michigan reports that the entire central alarm monitoring system should have a complete validation at a minimum of every 3 years to demonstrate that the system is operating properly. Any repairs to the monitoring system should be evaluated to determine if a full or partial validation to the system is needed. This would also include software upgrades and hardware changes.
ADDENDA Jan. 30, 2002
- A blood banker in Illinois wrote that it is his hospital's practice to contract with the vendor to validate the entire system, hardware and software, twice yearly. Alarm tests are performed quarterly (e.g. fire and ice testing) on each monitored unit. Validation at this frequency has prevented system failures by detecting them prior to a true issue developing. He adds that the Rees System (12 bit) has a resolution of 0.1 degrees C, not 0.2 degrees, as suggested by the responding blood banker in reply #1 above. Early versions were 8 bit and had a resolution of 0.4 degrees C. The Rees probes can be calibrated to the same tolerance as any NIST traceable thermometer.
Submit comments to the e-Network Forum at enetworkforum@cbbsweb.org
Ira A. Shulman, MD
CBBS e-Network Forum Senior Editor & Moderator
W. Tait Stevens, MD
CBBS e-Network Forum Editor & Moderator
Elizabeth M. St. Lezin, MD
CBBS e-Network Forum Associate Editor & Moderator
The e-Network Forum is supported in part by the California Blood Bank Society (CBBS) and the American Red Cross Blood Services (ARCBS) and endorses collegial discussion among blood banking and transfusion medicine professionals. However, neither the CBBS nor the ARCBS in any way endorse the specific views and opinions expressed in the forum. The forum is not intended as a substitute for medical or legal advice and the content should not be relied upon for any medical or legal purposes. Readers should make their own determinations as to: (i) what constitutes appropriate medical, technical, and administrative practices, and (ii) how best to comply with laws and regulations relevant to their questions. For the latter, they should consider consulting, as to any medical matters, a qualified physician, and, as to any legal matters, an attorney familiar with related state and federal laws. The user of the forum, by accessing same, assumes all risks arising out of such use and releases CBBS and their respective members, directors, officers and agents from and against any loss, damage, claim or liability arising out of such use of the Forum.